Influence of Heterogeneous Human Body on Midfield Wireless Power Transfer for Implantable Heart Devices

Influence of Heterogeneous Human Body on Midfield Wireless Power Transfer for Implantable Heart Devices

Abstract
DOI: 10.1109/TMTT.2023.3308628   IEEEXplore: https://ieeexplore.ieee.org/document/10242125
Device
Simulated midfield wireless power transfer (WPT) device assessed for optimal frequency for human tissue applications with complex tissue models.
Spectrum
1–2.5 GHz
Novelty
More than 30 multilayer tissue models based on real CT images of humans were utilized in the simulations to determine the optimal frequency and calculate the coupling parameter and the SAR efficiency.
Application
Cardiac pacemakers and other active implantable medical devices (AIMDs).
Performance
The simulations were validated with measurements with good agreement using a variety of non-human phantoms to demonstrate that the optimal frequency varied significantly with tissue composition and thickness.